Operation using the data setting device
Azbil Corporation
6-24
Model MTG11A/18A, MTG11B/18B, MTG14C
Setting engineering units: UNITS key
Introduction
The instantaneous flow rate value measured by the converter can be set so that it is
displayed in engineering units according to the control process used.
This setting is applied to both display screens of the display panel of the converter and
the SFC.
The engineering units that can be set are as follows.
When mass flow rate units are set, the specific gravity can be set.
Procedure
Use the following procedure to set engineering units.
Volume flow rate units
Mass flow rate units
m^3/h, GPH, l/h, cc/h,
m^3/min, GPM, l/min, cm^3/min,
m^3/day, GPD, kGPD, BPD,
m^3/s
GPH=gals/hr, GPM=gals/min,
GPD=gals/day, kGPD=1000XGPD,
BPD=barrels/day
kg/h, lb/h, kg/min, lb/min, kg/s,
lb/s, t/s, t/min, t/s, g/h,
g/min, g/s
Step
Key
Procedure
SFC screen
1
Confirm that the SFC is set to
“READY”.
If it is not, press the
key to set it
to “READY”
2
Press the
key.
Result:
• The currently set engineering unit
appears as shown on the right.
CLR
( No )
CLR
( No )
SFCM00006009D
D
UNITS
D
UNITS
SFCM00006039D
CM2-MTG300-2001.book 24 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...