![Azbil MagneW Neo PLUS Скачать руководство пользователя страница 253](http://html1.mh-extra.com/html/azbil/magnew-neo-plus/magnew-neo-plus_user-manual_3042666253.webp)
Azbil Corporation
Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C
6-69
5
Display shown to the right appears.
Press the
key and select the
value to be calibrated.
The value to gain calibrate for two
wired magnetic flowmeter are the
following three: 0 m/s, 2.5 m/s, and
10 m/s. Value 2.5 m/s for two
wired magnetic flowmeter will be
shown 1.2 m/s in SFC. When
calibrating to 2.5 m/s select 1.2 m/
s.
6
For example, when calibrating 2.5
m/s, displays screen as shown to
the right. For smart calibrator
model MGZ14, enter 2.5 m/s.
Then, press
key. Gain
calibration starts. When gain
calibration completes, display will
return back to the screen to select
the gain calibration value.
7
Follow the same procedure for
calibrating 0 m/s and 10 m/s.
8
When gain calibration is
completed, press the
key and
the display as shown on the right
appears.
9
Press the
key.
Exits from gain calibration
function.
10
Press the
key and the screen
returns to the screen as in step 1.
Step
Key
Procedure
SFC screen
MENU I
ITEM
DE CONF.
MENU I
ITEM
DE CONF.
SFCM00006125D
SFCM00006126D
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
SFCM00006126D
SFCM00006126D
SFCM00006128D
MENU I
ITEM
DE CONF.
MENU I
ITEM
DE CONF.
SFCM00006127D
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
SFCM00006128D
SFCM00006129D
CLR
( No )
CLR
( No )
SFCM00006009D
CM2-MTG300-2001.book 69 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...