Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C
5-35
6
Press the key to move the cursor to the position
under *.
Pressing the MODE key reboots the converter,
returns to MEASURING MODE, and the
communication system changes and saved.
m
CAUTION
You have only ten minutes to return to MEASURING MODE to save the new value before
the system resets it to the previously saved value.
Step
Procedure
Screen
20.0 %
* COM SELECT
HART
8.8.8.8.8.8.8.8
RATE
TOTAL
%
SELF CHECK
MODE
20.0 %
01.94 m
3
/h
WP0
00069401