![Azbil MagneW Neo PLUS Скачать руководство пользователя страница 168](http://html1.mh-extra.com/html/azbil/magnew-neo-plus/magnew-neo-plus_user-manual_3042666168.webp)
Operation using the data setting device
Azbil Corporation
5-62
Model MTG11A/18A, MTG11B/18B, MTG14C
5-7-1 : Configuration of OUTPUT CHECK MODE
Introduction
OUTPUT CHECK MODE has the following setting and adjustment items.
For details on the function of items, see the following pages.
Item
Content
Screen
OUTPUT
CHECK
W/CALIBRATOR
Perform loop checks by using a
calibrator (Model MGZ).
OUTPUT
CHECK
I.OUT
Outputs a fixed value of analog
current output to perform loop
checks.
OUTPUT
CHECK
P.OUT
Outputs a fixed value of pulse output
to perform loop checks.
OUTPUT
CHECK
ST.OUT
Outputs a fixed value of contact
output to perform loop checks.
OUTPUT
CHECK
EX
Outputs a fixed value of excitation
current.
This value was calibrated in the
factory.
DO NOT configure this value.
20.0 %
_> OUTPUT CHECK
W/CALIBRATOR
12.3 %
> OUTPUT CHECK
I.OUT
100%
12.3 %
> OUTPUT CHECK
P.OUT 100%
12.3 %
> OUTPUT CHECK
ST.OUT CLOSE
12.3 %
> OUTPUT CHECK
EX
EX1
CM2-MTG300-2001.book 62 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...