Operation using the data setting device
Azbil Corporation
5-68
Model MTG11A/18A, MTG11B/18B, MTG14C
5-7-6 : Configuration of CALIBRATION MODE
Introduction
CALIBRATION MODE has the following setting and adjustment items:
Configuration of CALIBRATION MODE requires a dedicated calibrator.
Wrong operation may hinder accurate measurements of the flow rate. To operate in
this mode, contact an Azbil Corp. representative.
Item
Content
Screen
CAL EX
LOW
3.5 mA
Adjusts 3.5 mA excitation
current.
CAL EX
4.9 mA
Adjusts 4.9 mA excitation
current.
CAL EX
7.0 mA
Adjusts 7.0 mA excitation
current.
CAL EX
11.9 mA
Adjusts 11.9 mA excitation
current.
CAL EX
14.0 mA
Adjusts 14.0 mA excitation
current.
CAL I.OUT
LOW
4.000 mA
Adjusts 4 mA analog current
output.
CAL I.OUT
HIGH
20.00 mA
Adjusts 20 mA analog current
output.
CAL P.OUT
FREQ
90 Hz
Adjusts 90 Hz pulse output.
12.3 %
> CAL EX
OFF
LOW
3.5
mA
12.3 %
> CAL EX
OFF
4.9
mA
12.3 %
> CAL EX
OFF
7.0
mA
12.3 %
> CAL EX
OFF
11.9
mA
12.3 %
> CAL EX
OFF
14.0
mA
12.3 %
> CAL I.OUT OFF
LOW
4.000
mA
12.3 %
> CAL I.OUT OFF
HIGH 20.000
mA
12.3 %
> CAL P.OUT OFF
FREQ
90
Hz
CM2-MTG300-2001.book 68 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...