Maintenance and troubleshooting
Azbil Corporation
8-8
Model MTG11A/18A, MTG11B/18B, MTG14C
Troubles during operation
Troubleshooting
When a problem occurs during operation, perform the following procedures.
1. Search for the symptom of the trouble in the table on this page. If found, perform
the steps indicated in the table.
2. If communication is possible, read the error message and self-diagnosis results.
Perform referring to “Error messages and troubleshooting”.
3. If the problem can not be solved, it is possible that the device has been damaged.
Contact an Azbil Corp. representative.
Trouble
Check point and troubleshooting
Output fluctuates
excessively beyond the
estimated flow rate range
• Check if the instrument is grounded correctly.
• Check if the damping time constant is set correctly.
• Clean the electrodes.
• The analog output may fluctuate due to flow noise,
which is generated by the process fluid flow. In
such a case, connect the upstream grounding ring to
the downstream grounding ring by a wire. The out-
put fluctuation may be reduced.
• A circuit in some PLC may affect the flow measure-
ment and the analog output may fluctuate. In this
case, make sure that the both PLC and the MTG
flowmeter are properly grounded. Proper grounding
solves the fluctuation problem.
Output exceeds 100%.
• Check if the range is set correctly.
• Check if zero is adjusted correctly.
Output remains 0%
• Check if the signal line is connected correctly.
• Check the upstream and downstream valves.
• Check if the range is set correctly.
• Check whether the instrument is set to the constant
current mode.
• Press the CLR (clear) key on SFC to cancel this
mode.
• Check if the flow rate is not within the set range of
low-flow cutoff.
• Check if the flow rate is not reversed (negative flow
rate).
• Check the detector for unfilled condition, too low
conductivity, excessive noise, etc.
Output has burnt out
• Take measures referring to “Error messages and
measures”.
Pulse output is too large or
too small for the calorific
value
• Is the pulse setting (weight and width) correct?
• Is the output from the main unit correct?
• Is a pulse counter of appropriate specifications
used?
• Is the dropout value correctly set between 0 and
10%.
CM2-MTG300-2001.book 8 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...