Operation using the data setting device
Azbil Corporation
5-16
Model MTG11A/18A, MTG11B/18B, MTG14C
5-5-3 : Setting of built-in counter reset value
Set a start value of the built-in counter. The scale of this value is considered as the
weight of the pulse. Carry out the built-in counter reset in Section 5.4.5 to start
totalization from any totalized value.
Set range:
00000000 - 99999999
Default:
00000000
Set an internal counter reset value in accordance with the following procedure:
Step
Procedure
Screen
1
The screen at the right is a screen display example
in MEASURING MODE (measurement state).
Press the MODE key.
2
The OPERATOR'S MODE screen appears for
approx. two seconds and then the damping setting
screen appears.
3
Press the key twice to display the screen.
4
Press the key to move the cursor to the position
under a target value to be changed.
5
Press the or key to set a desired value.
6
Press the key to move the cursor to the position
under *.
Press the MDOE key to return to the
MEASURING MODE and to save data.
m
CAUTION
You have only ten minutes to return to MEASURING MODE to save the new value before
the system resets it to the previously saved value.
20.0 %
01.94 m
3
/h
WP0
00069401
20.0 %
* OPERATOR'S
MODE
20.0 %
* DAMPING
005.0 s
20.0 %
* CNT-RESET VALUE
00044444
20.0 %
* CNT-RESET VALUE
00000000
20.0 %
* CNT-RESET VALUE
00005000
20.0 %
* CNT-RESET VALUE
00005000
CM2-MTG300-2001.book 16 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...