Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C
5-21
5-5-7 : Setting Electrode Status Diagnostic function
Overview of electrode status diagnostic function
Electrode status diagnostic function detects the condition of the empty pipe or the
scale on the electrodes. The Electrode status diagnostic function makes the analog
output and pulse output to the values as selected in the below “Electrode status output
mode” table.
The display alternately shows the output values selected and “EMPTY OR SCALE
ON ELECTRODE”.
“Electrode status output mode” table
Output/Display
Parameter selection in the “Electrode status output mode”
OFF
ZERO
HOLD
Analog 4 – 20mA
output
Output values as the
meter measures.
Analog output is
fixed to 0% (4mA).
Analog output is
held at its last good
value.
Pulse output
Output values as the
meter measures.
Pulse output is fixed
to 0 (does not
generate pulses).
Pulse output is held
at its present state.
Display
Display the value as
it measures.
Flashes the message
0% and “Empty or
scale on electrode”
alternately (when %
flow rate is
specified for the
main display).
Flashes the message
0.000 RATE and
“Empty or scale on
electrode”
alternately (when
actual flow rate is
specified for the
main display).
Flashes the message
XXXXXXXX
(totalized value at
setup) and “Empty
or scale on
electrode”
alternately (when
totalized value is
specified for the
main display).
Flashes the values at
its last good values
and a message of
“Empty or scale on
electrode”
alternately.
CM2-MTG300-2001.book 21 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...