Instrument installation
Model MTG11A/18A, MTG11B/18B, MTG14C
2-13
Flange shape
The flanges used should be such that the area of contact with the gasket is maximized,
as shown in Figure 2-10.
Figure 2-10 Flange shape
~Note
•
Before installing the detector be sure to flush out any foreign matter
that may be present inside the detector. Residual foreign matter could
cause output fluctuations.
•
Do not touch the electrodes by your hand or waste cloth with oil. It
could cause output fluctuations.
•
Install the detector in accordance with the flow direction mark on the
detector with the direction of the liquid flow. Misalignment could
result in a negative output.
Acceptable
X Unacceptable
Flange
Welding
Welding
Pipe
(The liquid could
leak because of
the small area
of contact with
the gasket.)