User Manual
116
i Series / iX Series
150 VDC Range Volt Zero:
Press the PROGRAM key. Select VOLT RANGE and program the 150 VDC range with the
shuttle. Go to the OUTPUT CAL screen and adjust the VOLT ZERO for 0.0 0.005 volts DC on
the output.
150 VDC Range + Full-scale:
P120.0 volts. Go to the OUTPUT CALIBRATION screen and adjust the VOLT FS
parameter for an output voltage of +120.0 ±0.05 volts DC.
150 VDC Range - Full-scale:
Program -120.0 volts. Go to the OUTPUT CALIBRATION screen and adjust the VOLT FS
parameter for an output voltage of -120.0 ±0.05 volts DC.
For a 9003iX or 15003iX 3-Phase power system with one controller, repeat the preceding steps
for the Phase B and C outputs. Press the PHASE key to select the output to be calibrated.
Monitor the output of the respective phase by moving the HI input of the Digital Multimeter. The
LO input should remain connected to the common LO of the sense connector, TB3.
The following Table is a summary of the preceding calibration steps.
Program the following values in the table and make the adjustments in the OUTPUT
CALIBRATION screen. Select the phase to be calibrated by pressing the PHASE key.
TITLE
PROGRAM VALUES
CALIBRATION VALUE
ADJUST TO
150 VAC range DC Zero
150 range, 0.0 V
VOLT ZERO
0 ± 5 mv DC
150 VAC range Volt FS
120.0 V, 60 Hz
VOLT FS
120 0.05 VAC
150 VAC range Volt Hi Freq
120.0 V, 400 Hz
V HI FREQ
120 0.05 VAC
Repeat the adjustments at 60 and 400 Hz until the output is within ± 0.05 volts
300 VAC range DC Zero
270 range, 0.0 V
VOLT ZERO
0 ± 5 mv DC
300 VAC range Volt FS
240.0 V, 60 Hz
VOLT FS
240 0.05 VAC
300 VAC range Volt Hi Freq
240.0 V, 400 Hz
VOLT HI FREQ
240 0.05 VAC
Repeat the adjustments at 60 and 400 Hz until the output is within 0.05 volts
150 VDC range DC Zero
135 range, 0.0 V
VOLT ZERO
0 ± 5 mv DC
150 VDC range Volt +FS
+120.0 V
VOLT FS
+120 0.05 VDC
150 VDC range Volt -FS
-120.0 V
VOLT FS
-120 0.05 VDC
300 VDC range DC Zero
270 range 0.0 V
VOLT ZERO
0 ± 5 mv DC
300 VDC range Volt
+240.0 V
VOLT FS
+240 0.05 VDC
300 VDC range Volt -FS
-240.0 V
VOLT FS
-240 0.05 VDC
Table 6-1: Output Calibration Table
For a multi-phase power system that uses one controller, 9003iX or 15003iX, repeat paragraph
6.4 for each phase. Move the external multimeter to the phase being calibrated. Make sure that
the phase being calibrated is programmed to the correct voltage. While viewing the calibration
screen, press the PHASE key to select the respective key.
Содержание 10001i
Страница 2: ......
Страница 3: ......
Страница 6: ...ii This page intentionally left blank...
Страница 25: ...User Manual i Series iX Series 11 Parameter Specification Shock Designed to meet NSTA 1A transportation levels...
Страница 38: ...User Manual 24 i Series iX Series Figure 3 5 Rear Panel View for the 3001i 3001iX...
Страница 39: ...User Manual i Series iX Series 25 Figure 3 6 Rear Panel View for the 5001i 5001iX...
Страница 42: ...User Manual 28 i Series iX Series Figure 3 7 Connection For Single Power Source 5001iX i 3001iX i...
Страница 43: ...User Manual i Series iX Series 29 Figure 3 8 Functional Test Setup...
Страница 44: ...User Manual 30 i Series iX Series Figure 3 9 Single Phase 10000 VA System 10001iX i...
Страница 45: ...User Manual i Series iX Series 31 Figure 3 10 Three Phase 15000 VA System 15003iX i LK Three Controllers...
Страница 46: ...User Manual 32 i Series iX Series Figure 3 11 Single Phase 15000 VA System 15001iX i...
Страница 47: ...User Manual i Series iX Series 33 Figure 3 12 Three Phase 15000 VA system 15003iX i One Controller...
Страница 48: ...User Manual 34 i Series iX Series Figure 3 13 Connection With MODE Option...
Страница 49: ...User Manual i Series iX Series 35 Figure 3 14 Two Phase 10000 VA System 10002i LK Two Controllers...
Страница 50: ...User Manual 36 i Series iX Series Figure 3 15 Three Phase 9000 VA System 9003iX i One Controller...
Страница 118: ...User Manual 104 i Series iX Series Figure 5 2 Power Source Module Block Diagram...
Страница 121: ...User Manual i Series iX Series 107 Figure 5 3 5001i Internal Layout...
Страница 122: ...User Manual 108 i Series iX Series Figure 5 4 Logic Board LED s...
Страница 124: ...User Manual 110 i Series iX Series Figure 5 5 AC Power Stage Layout...
Страница 125: ...User Manual i Series iX Series 111 Figure 5 6 AC Control Logic Block Diagram...
Страница 132: ...User Manual 118 i Series iX Series Figure 6 2 Test Equipment Hook up for Measurement Calibration...
Страница 138: ...User Manual 124 i Series iX Series Figure 6 3 Adjustment Location...
Страница 152: ...User Manual 138 i Series iX Series Figure 9 4 Voltage Modulation...
Страница 219: ...User Manual i Series iX Series 205 Figure 9 36 Example Connection With 5001iX and EOS 1...
Страница 220: ...User Manual 206 i Series iX Series Figure 9 37 Example Connection With Compliance Test System and EOS 1...
Страница 221: ...User Manual i Series iX Series 207 Figure 9 38 15003iX CTS EOS3 LR3...
Страница 222: ...User Manual 208 i Series iX Series Figure 9 39 15003iX 3 EOS3...
Страница 226: ...User Manual 212 i Series iX Series Figure 9 40 EOS3 Location of 70 80 Taps for each phase Lug 3 70 Lug 5 80...
Страница 233: ...User Manual i Series iX Series 219 Figure 9 41 Example Connection With MODE iX...
Страница 240: ...User Manual 226 i Series iX Series Figure 9 42 Example Connections With OMNI 1 18i...
Страница 241: ...User Manual i Series iX Series 227 Figure 9 43 Example Connections With OMNI 3 18i...
Страница 242: ...User Manual 228 i Series iX Series Figure 9 44 Schematic Showing OMNI 1 37i and1 37iJ Connected to 5001iX System...
Страница 243: ...User Manual i Series iX Series 229 Figure 9 45 Schematic Showing OMNI 3 37i Connected to 30003iX System...
Страница 249: ...User Manual i Series iX Series 235 9 7 4 Mechanical Dimensions Figure 9 49 XLS Module Dimensions...