10
Field of view.
%
Humidity limitation: relative humidity between
30 and 85 % in operating environment.
Atmospheric pressure limitation: between 80 and
109 kPa in operating environment.
Electrical Safety Type BF Applied Part.
Packaging level ensuring sterility.
UL Recognized Component Mark for Canada and
the United States.
Medical device.
Global Trade Item Number.
Country of manufacturer.
INMETRO Certificate Medical Electrical Equipment
ABNT NBR IEC 60601-1, ABNT NBR IEC 60601-1-2,
ABNT NBR IEC 60601-2-18.
Applicable for Ambu® aScope™ 4 Broncho and
Ambu® aView™ monitor.
Do not use if the product sterilisation barrier
or its packaging is damaged.
4. Use of aScope 4 Broncho Sampler Set
Numbers in gray circles below refer to illustrations on page 2.
4.1. Preparation and inspection of aScope 4 Broncho Sampler Set
Visual inspection of the endoscope 1
1. Check that the pouch seal is intact.
1a
2. Make sure to remove the protective elements from the handle and from the
insertion cord.
1b
3. Check that there are no impurities or damage on the product such as rough surfaces,
sharp edges or protrusions which may harm the patient.
1c
Refer to the Ambu displaying unit
Instruction for use
for preparation and inspection
of the Ambu displaying unit 2
Inspection of the image
1. Plug in the endoscope cable connector into the corresponding connector on the
compatible displaying unit. Please ensure the colours are identical and be careful to align
the arrows.
3
2. Verify that a live video image appears on the screen by pointing the distal tip of the
endoscope towards an object, e.g. the palm of your hand.
4
3. Adjust the image preferences on the displaying unit if necessary (please refer to the
displaying unit
Instruction for use
).
4. If the object cannot be seen clearly, wipe the lens at the distal tip using a sterile cloth.
Preparation of aScope 4 Broncho Sampler Set
1. Carefully slide the control lever forwards and backwards to bend the bending section as much
as possible. Then slide the control lever slowly to its neutral position. Confirm that the bending
section functions smoothly and correctly and returns to a neutral position.
5a
Содержание aScope 4 Broncho Large
Страница 17: ...BG 17 5 4 6 6 4 1 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Ambu Ambu 24...
Страница 18: ...18 25 26 27 28 1 2 3 4 ETT 5 6 7 8 9 10 85 kPa 638 mmHg 11 1 5...
Страница 21: ...BG 21 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Luer Lock B 15 16 17 18...
Страница 28: ...28 6 2 2...
Страница 29: ...BG 29 aScope 4 Broncho aScope 4 Broncho...
Страница 69: ...EL 69 5 4 6 6 4 1 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Ambu Ambu 24...
Страница 70: ...70 25 26 27 28 1 2 3 4 5 6 7 8 9 10 85 kPa 638 mmHg 11 1 5 1 6...
Страница 73: ...EL 73 A aScope 4 Broncho 1 2 3 4 5 6 7 8 9 LED 10 11 12 13 14 Luer Lock B 15 flip top 16...
Страница 80: ...80 6 2 2...
Страница 81: ...EL 81 aScope 4 Broncho aScope 4 Broncho...
Страница 170: ...170 14 15 16 17 18 19 20 21 22 23 Ambu Ambu 24 25 26 27 28 1 2 3 4 ETT 5 6 7 8 9 10 85 kPa 638 mmHg 11...
Страница 179: ...JA 179 30 ml aScope 4 Broncho ETO aScope 4 Broncho C F 10 40 50 104 30 85 kPa 80 109 1 2 3 6 2 2...
Страница 180: ...180 aScope 4 Broncho aScope 4 Broncho...
Страница 269: ...RU 269 5 4 6 6 4 1 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Ambu Ambu...
Страница 270: ...270 24 25 26 27 28 1 2 3 4 ETT 5 6 7 8 9 10 85 638 11 1 5...
Страница 273: ...RU 273 1 2 3 4 5 6 7 8 9 10 11 12 13 14 B 15 16 17 18 C...
Страница 280: ...280 6 2 2...
Страница 281: ...RU 281 aScope 4 Broncho aScope 4 Broncho...
Страница 339: ...ZH 339 6 2 2 aScope 4 Broncho aScope 4 Broncho...
Страница 340: ...Ambu A S Baltorpbakken 13 DK 2750 Ballerup Denmark T 45 72 25 20 00 ambu com 492 4770 00 V03 2021 07 TCC 11078...