E 740
Greasing molecular and
turbomolecular pumps
G
B
01240
-
E
d
it
io
n
04
-
J
ui
n
08
Bearing opposite
the pumping cell
Introduce the syringe equipped with its needle into
the bearing and remove the red jumper.
Always position the syringe needle between two balls of the
bearing so as not to damage the bearing.
Distribute the dose of grease in 2 diametrically
opposed points, until the syringe plunger comes to
a stop against the black jumper.
Bearing on pumping
cell side
3
Introduce the lubrication syringe needle
into the drilled screw located at the center
of the rotor until it comes to a stop against
the screw head.
Keep the syringe pressed down to the bottom of
its housing throughout the operation.
Remove the black jumper from the syringe and
introduce the grease until the plunger comes to a stop.
Remove the syringe.
Reassemble the adjustment sleeve (taking care with the direction),
the spring and the end cap with its o-ring.
The relubrication operation is complete.
Execute the pump running-in.
5/5