C 200
Starting up / Switching off the leak
detector
3/3
CYCLE IN PROGRESS
MODE:->
Normal
...
...
GB
02460
-
E
d
it
io
n
02
-
J
une
08
Recommended
procedure
To protect the internal vacuum components of the leak detector against
dust or any kind of contamination, it is recommended to keep its inlet
blanked-off and under vacuum. Please proceed as follows before
stopping the leak detector:
Place the blank-off flange or close the test chamber.
Start the test cycle.
Wait until the leak detector reaches the more sensitive test mode.
Make sure that the inlet vent is OFF.
Stop the test cycle.
Stop the leak detector.