G 400
Long distance sniffer probe user manual
GB
01962
-
E
d
it
io
n
03
-
J
une
08
Filter exchange
(continued)
When you change the filter, we recommend cleaning the needle and
the nipple (sniffer probe with rigid nipple only):
Take out the needle with needlenose pliers.
Clean delicately the needle with alcohol and a lint-free cloth.
Clean the rigid nipple with alcohol and compressed dry air.
Put back the needle.
Put back the filter with its screw: torque < 1 N.m.
Needle replacement
With this kind of sniffer probe, it is normally not necessary to change
the needle.
For every needle replacement in a sniffer probe, put a new needle
with the same diameter as the old needle.
3 methods can be used for the new needle adjustment :
1
flowmeter use
2
standard sniffer probe use
3
old needle use
Recommended methods
1
Flowmeter use in order to
measure the flow in the sniffer
probe
2
Standard sniffer probe use as
reference
Assembly to do:
Helium leak
detector
P < 10 mbar
or
Flowmeter
0/100 sccm
Pump
Sniffer probe
connected to
the assembly
This method requires keeping a new
sniffer probe as a standard probe.
Do an auto-calibration in sniffing
mode with the standard sniffer
probe.
Do a measure of the He in the air
with the standard sniffer probe.
Cut the new needle to 85 mm/3.35 inch. Prepare it (see below “
3
Use of
the old needle as a reference“, picture
3
) and put it correctly in the nozzle
(see “Needle exchange” §).
5/8
Содержание ASM 142
Страница 1: ...ASM 142 142 D ASM GRAPH GRAPH D GRAPH D HELIUM LEAK DETECTOR User s Manual ...
Страница 98: ......
Страница 106: ......
Страница 111: ......
Страница 119: ......
Страница 124: ......
Страница 129: ......
Страница 140: ......
Страница 144: ......
Страница 166: ......
Страница 315: ...G 201 Wiring diagrams ASM 142 D index 1 2 ...