A 400
Testing methods
GB
000
0
8
-
E
d
itio
n
05
-
F
ebr
uar
y
02
Helium concentration
and signal displayed
In accordance to the He concentration rate in the gas used for the
leak detection, the signal displayed will change.
Example: signal displayed with a 1x10
-7
mbar.l/s calibrated leak
(with 100 % He) connected to the detector inlet.
% He in the gas used
100 %
10 %
1 %
Signal displayed
on the leak detector
1x10
-7
mbar.l/s
1x10
-8
mbar.l/s
1x10
-9
mbar.l/s
Spray method
(inboard testing)
This involves removing air from the test part, connecting it to the
analyzer and then spraying helium over the outer surface.
Global test
Spray test
or
part
spray probe
He
part
He
detector
detector
The part is placed under a
cover, into which helium is
injected.
The leak cannot be located.
Potential leaking areas are
sprayed with helium.
The leak can be located.
The detector measures the flow of helium penetrating the part.
Response time
When spraying starts, the leak signal is not displayed instantaneously
on the analyzer:
there is a response time which depends on the volume V being tested
and the helium pumping speed S of the system at the opening of the
part, according to the following relation:
T = V/S
(T in seconds, V in litres, S in l/s)
T is the time required for the signal to reach 63 % of the final value.
2/4
Содержание ASM 142
Страница 1: ...ASM 142 142 D ASM GRAPH GRAPH D GRAPH D HELIUM LEAK DETECTOR User s Manual ...
Страница 98: ......
Страница 106: ......
Страница 111: ......
Страница 119: ......
Страница 124: ......
Страница 129: ......
Страница 140: ......
Страница 144: ......
Страница 166: ......
Страница 315: ...G 201 Wiring diagrams ASM 142 D index 1 2 ...