C 211
Operation of the leak detector
4/4
READY FOR CYCLE
INLET: ...
SNIFFING MODE
READY FOR CYCLE
INLET ...
GB
01362
-
E
d
it
io
n
02
-
S
e
p
te
m
b
e
r
04
Sniffing test
Starting a sniffing test While the leak detector is in
stand-by mode, connect the sniffer
probe (accessory to be purchased
separately
A 700) to the sniffer port
of the leak detector.
STDBY
SNIFFING MODE
PLEASE WAIT
...
The sniffing mode message
appears on the alphanumeric
display.
The sniffing test mode is operational.
Leak value display
Ending a sniffing test
Displays are the same in hard vacuum and sniffing test modes.
Please refer to hard vacuum test mode for the displays.
STDBY
Sniffer probe clogged
reject point
Accessories
A 700
Advice Block the sniffer probe end from time to time with a finger to check
that the helium signal goes down. If not, the probe may be clogged.
Sniffer probe clogged