Factory configuration of
the leak detector parameters
Parameters
configuration
The following list indicates the factory configuration of the leak detector
parameters.
When the leak detector is switched off, all set parameters are
memorized and values are kept for the next start-up.
We advice you to note in the “Customer modification“ column, the
parameter values modified for your application.
Parameters
Configuration
Factory
Customer
modification
Test mode
Normal
C 210
Sniffer probe clogged threshold
1.10
- 7
mbar.l/s
C 211
Autocalibration
Activation
Auto-checking every …
Auto-checking every ...
On
Operator
10 hours
100 cycles
C 302
External calibrated leak in test
External calibrated leak in sniffing
1.10
- 7
mbar.l/s
5.10
- 6
mbar.l/s
C 303
GL correction
GL correction value
Sniffing correction
Sniffing correction value
Off
1
Off
1
C 304
Calibrated leak parameters
See calibration
certificate of
the internal
calibrated leak
delivered with
the detector
.
C 305
Password
5555
C 130
User interface level
2
C 120
Audio alarm
Voice synthesis
Reject point in hard vacuum
Reject point in sniffing
3
4
1.10
- 8
mbar.l/s
1.10
- 6
mbar.l/s
C 520
Detector Inlet vent
Inlet vent activation
Delay
Open timer use
Timer
Off
Operator
0
No
00:10
C 500
Depollution
Depollution reject point
On
1.10
- 5
mbar.l/s
C 560
Содержание ASM 142
Страница 1: ...ASM 142 142 D ASM GRAPH GRAPH D GRAPH D HELIUM LEAK DETECTOR User s Manual ...
Страница 98: ......
Страница 106: ......
Страница 111: ......
Страница 119: ......
Страница 124: ......
Страница 129: ......
Страница 140: ......
Страница 144: ......
Страница 166: ......
Страница 315: ...G 201 Wiring diagrams ASM 142 D index 1 2 ...