G 400
Long distance sniffer probe user manual
Description
P/N
Adaptor for DN 16 calibrated leak
A 006
Adaptor for DN 25 calibrated leak
A 007
GB
01962
-
E
d
it
io
n
03
-
J
une
08
"Sniffer probe clogged"
message
A “Sniffer probe clogged” message could appear on the control panel
LCD display or be announced by the digital voice: the leak detector
compares the helium signal read on the detector to the sniffer probe
clogged reject threshold.
10
-7
mbar.l/s
Sniffer probe clogged message
No message
10
-5
mbar.l/s
Sniffer probe clogged
reject threshold value
Value of the helium signal
read on the detector
During the needle adjustment, this message could appear without the
snifffer probe necessarily being clogged: this is why the needle length
is so important.
For more details, please consult the user’s manual delivered with your
leak detector.
Advice:
Block the sniffer probe end from time to time with a finger to check
that the helium signal goes down. If not, the sniffer probe may be
clogged.
Adaptor for calibrated
leak
Reference part number
F 1000
Special adaptors for calibrated
leaks have been designed to ensure
a good connection and repetitive
and reliable calibration with a
sniffer probe.
Adaptor
Calibrated
leak
Sniffer
probe
With the adaptor for calibrated leaks use:
calibrated leak value
Value read on the leak detector
=
{
+
value of the helium in the air
8/8
Содержание ASM 142
Страница 1: ...ASM 142 142 D ASM GRAPH GRAPH D GRAPH D HELIUM LEAK DETECTOR User s Manual ...
Страница 98: ......
Страница 106: ......
Страница 111: ......
Страница 119: ......
Страница 124: ......
Страница 129: ......
Страница 140: ......
Страница 144: ......
Страница 166: ......
Страница 315: ...G 201 Wiring diagrams ASM 142 D index 1 2 ...