B 110
Unpacking - Storage - Transportation
GB
00215
-
E
d
it
io
n
04
-
J
une
08
Fixed mounting
The ASM l42 can be fixed on a support: a threaded hole is available
on the bottom of the unit to place a screw to keep it secure to a
working surface such as a bench (except ASM Graph D+).
M8 screw
(30 + x) mm< L < (45 + x) mm
L
x
Transport
Before expedition, we advise you to drain the primary pump to avoid
any oil split (only ASM l42 - ASM l42 S).
Don’t move leak detector when AMP pump is running.
We advise for any transport to use the original packaging and to
wedge the detector carefully into the box.
Precautionary
measures for the leak
detector installation
Only for ASM Graph D+
The cart is equipped with 4 pivoting wheets with brake.
It is forbidden to place the cart on a sloping surface (> l %) without
take precautions for its translation stop.
4/4
Содержание ASM 142
Страница 1: ...ASM 142 142 D ASM GRAPH GRAPH D GRAPH D HELIUM LEAK DETECTOR User s Manual ...