G 400
Long distance sniffer probe user manual
GB
01962
-
E
d
it
io
n
03
-
J
une
08
Available spare parts
2
1 3
1
2
3
4
Designation
Reference
dia. 0.30 mm needle
J 001
1
dia. 0.35 mm needle
J 002
Rigid nozzle of 9 cm/3.54 inch (*)
J 003
2
Rigid nozzle of 30 cm/11.81 inch (*)
J 004
Flexible nozzle of 15 cm/5.90 inch (*)
J 005
Flexible nozzle of 45 cm/17.71 inch (*)
J 006
(*) (delivered with the suitable needle not cut)
Screw alone
J 007
3
Kit for sniffer probe with 5 filters, 2 O’rings and 2 screws
J 008
4
Reference part number
F 1000
Filter exchange
With the 2.5 Allen key, remove the screw at the end of the sniffer
probe (ref.
1
).
Take out the old filter. Put the new filter in the port (ref.
2
).
Hold this assembly straight up and screw on the nipple of the sniffer
probe (ref.
2
and
3
): torque < 1 N.m
Install the screw so that it is totally inside the nipple : you should see
a few threads (ref.
4
and
5
).
1
2
3
4/8
4
Screw inside
5
the nipple
Good position