C 306
1/1
AUTOCAL IN PROGRESS
ELECT. ZERO ADJUST.
G
B
01922
-
E
d
it
io
n
03
-
J
un
e
08
Adaptor for calibrated leak in sniffing
mode
An adaptor DN 16 or DN 25 for calibrated
leak has been designed for the calibration of
the detector with an external calibrated leak.
Accessories
A 700
How to use the
Adaptor
Sniffer probe
adaptor ?
3
2
1
4
5
Calibrated leak
1
Place the adaptor to your calibrated leak used for the calibration.
Start a
calibration
2
Place the sniffer probe in the calibration port.
3
Tighten the fixing srew. Follow the auto-calibration.
Calibration of the leak detector
C 300
4
Untighten the fixing screw.
5
Remove the sniffer probe of the calibration port. Follow the auto-
calibration.
Notes
- Waiting 10 s (mini) for the signal stabilization before reading of the
leak value.
- The leak value displayed on the LCD consider the He of the air.
Example : calibration with a leak of 2x10
-5
mbar.l/s
The value displayed is:
2x10
-5
+ 5x10
-6
= 2.5x10
-5
mbar.l/s