E 110
Access to the internal components
ASM 142 / ASM 142 S
d
Internal calibrated leak (
GB
002
3
8
-
E
d
itio
n
04
-
F
ebr
ua
ry
06
Access level 1
(continued)
Access to the internal
calibrated leak
Disconnect the calibrated leak temperature sensor.
Tighten the calibrated leak clips between two fingers and pull the
calibrated leak own
Temperature
sensor
Internal calibrated leak (*)
5
Access level 2
Access to the board
P0326 and P0330
(*) ASM 142 S in option
To be applied after access level 1 is completed.
Remove the two screws M5 and their washers
1
Open the removable plate
1
Disassemble the main block
Remove all the connections:
2 connections for analyzer cell
1
2 quick connectors (for valves)
2
2 screws M5 + washers
1
2