E 400
Basic maintenance of the analyzer cell
3
1
5
6
2
4
GB
00046
-
E
d
it
io
n
07
-
J
une
08
In case the analyzer cell is equipped with the optional metal seal:
g
Prepare the new special metal seal according to the figure
below or using the seal former,
F 110
g
The ends of the metal seal must only cross once (no twist). Check
that the ends cross near one of the six screws holes, one end on
either side of the hole. Place the metal seal on the cell body seal
seat.
g
Install the flange, taking care to lower it into the duct without
touching the sides.
g
Install the 6 screws.
Tighten all
the securing screws
5
g
Tighten the screws with their washers in the sequence shown
below to a torque of 0.7 m.daN for the special elastomer seal
(0.8 m.daN or 8 ft. lbs for the metal seal).
CAUTION
Don’t take off the
black sticker
6 screws tightening order
6/8
Содержание ASM 142
Страница 1: ...ASM 142 142 D ASM GRAPH GRAPH D GRAPH D HELIUM LEAK DETECTOR User s Manual ...
Страница 98: ......
Страница 106: ......
Страница 111: ......
Страница 119: ......
Страница 124: ......
Страница 129: ......
Страница 140: ......
Страница 144: ......
Страница 166: ......
Страница 315: ...G 201 Wiring diagrams ASM 142 D index 1 2 ...