A help for control panel utilization/access.
Operating principle of the control panel
C 110
Access to level
4
- Password
C 130
Setting and maintenance part presentation of the control panel
Access to parameters and parameters active
depending on authorization
C 120
Summary of screens
Complete displays list with access
way and associated sheet
C 140
Purpose of the cycle
end
It allows an automatic control of the roughing time and measure
time i. e. the end of cycle.
•
Cycle end: operator - Operator stops the cycle (
)
•
Cycle end: automatic - Cycle is stopped automatically.
Note: if Cycle end is automatic, the parameters of roughing and
measure times are automatic.
Activate / Deactivate
the Cycle end
Adjustment
operator
4x
C 110
Adjustment
automatic
yes
no
measure
Adjustment
Adjustment
roughing
•
Roughing
Roughing timer determines maximum authorized roughing time before
the tested part is considered as bad.
To consider the tested part as bad, it is necessary to activate the
Memo function.
Memo function
C 550
•
Measure
If Cycle end is activated (automatic), user should adjust the measure
timer i.e. the cycle end.
The measure timer corresponds to the time allowed for the leak
detector to remain in test. When it is reached, the measure flashes.
1/1
Содержание ASM 142
Страница 1: ...ASM 142 142 D ASM GRAPH GRAPH D GRAPH D HELIUM LEAK DETECTOR User s Manual ...
Страница 98: ......
Страница 106: ......
Страница 111: ......
Страница 119: ......
Страница 124: ......
Страница 129: ......
Страница 140: ......
Страница 144: ......
Страница 166: ......
Страница 315: ...G 201 Wiring diagrams ASM 142 D index 1 2 ...