Fault / information indicator and display
9/6
AUTOCAL. FAILURE
BACKGROUND TOO HIGH
PRESS NEXT
G
B
02465
-
E
d
it
io
n
03
-
S
ep
tem
ber
06
Fault and
information
Faults
Minor fault
At any time, the leak detector can display on the LCD clear
Information or Fault messages based on the analysis of the leak
detector status.
There are 3 basic types of faults: minor fault, major fault and critical
failure.
There are 2 basic types of information: user information and service
information.
The messages are displayed on a specific display by order of
importance:
1. critical failure
2. major fault and minor fault
3. user information and service information
3 fault types: minor fault, major fault and critical failure.
•
Warning:
- on the digital display alternatively the helium signal and “Er” are
shown.
- on the LCD, a “!” flashing at the right end of the 1
st
line.
LEAK RATE
COR
.....
!
.....
.....
.....
The digital voice advises the operator of the procedure to follow.
•
Result:
This will not stop the functions of the leak detector but can affect the
validity of the test result.
•
Message:
To read the messsage
NEXT
A clear message describes the error
on line 2 and 3. The most important
warning message is displayed on the
1
st
line.
Up to 3 messages may be displayed on the message display.
•
Duration: may be temporary or permanent
- temporary if the fault appears and then disappears without a
corrective action from the user
- permanent until the cause is erased by the user.
•
Remedy:
- Temporary: the indicator disappears and the warning message is
erased.
- Permanent: both indicator and message are memorized until the fault
is eliminated.
Содержание ASM 142
Страница 1: ...ASM 142 142 D ASM GRAPH GRAPH D GRAPH D HELIUM LEAK DETECTOR User s Manual ...
Страница 98: ......
Страница 106: ......
Страница 111: ......
Страница 119: ......
Страница 124: ......
Страница 129: ......
Страница 140: ......
Страница 144: ......
Страница 166: ......
Страница 315: ...G 201 Wiring diagrams ASM 142 D index 1 2 ...