1/2
C 305
Calibrated leak values programming
GB
00971
-
E
d
it
io
n
05
-
Ma
rc
h
04
A help for control panel utilization/access.
Operating principle of the control panel
C 110
Access to level
4
- Password
C 130
Setting and maintenance part presentation of the control panel
Access to parameters and parameters active
depending on authorization
C 120
Summary of screens
Complete displays list with access
way and associated sheet
C 140
Different types of
calibrated leaks
Before to start an internal calibration of the leak detector (with internal
or external calibrated leak), the parameters of the calibrated leak used
should be programmed by the operator.
The leak detector can be calibrated:
•
with an internal or external calibrated leak
•
with different gases (Hydrogen and Helium 3) if it is equipped with
the 3 masses option.
3 masses option
C 430
Gas
Internal calibration
External calibration
Helium 4
X
X
Helium 3
-
X
Hydrogen
-
X
Whatever the type of calibrated leak used, the parameters to program
are the same.