G 400
Long distance sniffer probe user manual
GB
01962
-
E
d
it
io
n
03
-
J
une
08
Use precautions
with the flexible
sniffer probe
The sniffer probe nipple can be bent if necessary but you should
respect a minimum radius of curvature (ref.
4
and
5
).
(continued)
Sniffer probe with a flexible
nipple of 15 cm / 5.9 inch
(part number SNCxE3Tx)
Sniffer probe with a flexible
nipple of 45 cm / 17.7 inch
(part number SNCxE4Tx)
4
E
5
R
E
R
E
E
R > 5 cm (2 inch) E > 2 cm (1 inch): do not twist/bend the E section
Flow adjustment
In order to adjust the flow inside the sniffer probe, it comes equipped
with 2 needle types: dia. 0.30 mm (0.11 inch) or 0.35 mm
(0.14 inch). This choice is done in factory and it is permanent.
How to identify the needle set in your snifffer probe?
The needle shape is different:
• Ø 0.30 mm needle
• Ø 0.35 mm needle
The sniffer probe nozzle is marked:
Sniffer probe equipped with a
dia. 0.30 mm needle has no mark
or is marked “0”.
Sniffer probe equipped with a
dia. 0.35 mm needle is marked “5”.