C 440
Control panel with graphic interface
8/14
G
B
02669
-
E
d
it
io
n
03
-
J
un
e
08
8
8
2
5
6
7
1
3
1
7
4
2
3
4
Mode
“Basic“
3
4
5
6
7
8
Mode “Advanced“
•
Curves to show/recorder selection with a simple press on the
square: inlet pressure (PRESSURE), helium signal (HELIUM) and
reject threshold (REJECT POINT).
- The reject threshold and the unit of measurements in this menu
can also be adjusted in the SET POINTS and OTHER menus. Any
modification will be applied to the 3 menus (SET POINT, OTHER
and graphic interface).
•
Display area setting of the inlet pressure and helium signal curves.
- If only one curve is selected, the display area applies to this
curve.
- If the both are selected, the display area applies to the helium
signal : the inlet pressure display area will be automatically
5 decades higher than the helium signal display area.
Modification validation or return to the graph.
Sampling rate between 2 measurements = 2 consecutive points of the
memorized curve.
Maximum recording time possible according to the sampling time set
for the memorized graph.
Time window displayed for the dynamic graph.
- “fast“ represents the maximum signal scrolling speed.
Adjustment of the contrast.
Contrast adjustment
From the setting menu, you can adjust the screen contrast.
Optimum brightness after 5mn switch on. This panel is delivered with
an optimum setting of the brightness.
Note:
The screen lifetime is the best with a contrast adjusted to 50 %.
•
Press on the contrast key
8
•
Adjust the contrast with the
and
keys
1
•
Validate the modifications
Содержание ASM 142
Страница 1: ...ASM 142 142 D ASM GRAPH GRAPH D GRAPH D HELIUM LEAK DETECTOR User s Manual ...
Страница 98: ......
Страница 106: ......
Страница 111: ......
Страница 119: ......
Страница 124: ......
Страница 129: ......
Страница 140: ......
Страница 144: ......
Страница 166: ......
Страница 315: ...G 201 Wiring diagrams ASM 142 D index 1 2 ...