C 580
Fault / information indicator and display
1/6
WARNING!
PRESS NEXT
...
...
...
CRITICAL FAILURE
pressure gauge fail.
SERVICE AND RESTART
G
B
02465
-
E
d
it
io
n
03
-
S
ep
tem
ber
06
Faults (cont.)
ASM 142 S and ASM 102 S special case:
•
If after the start, the background is lower than the sniffer probe
clogged point:
- The “AL” message appears permanently on the
digital display,
- a «!» flashing at the right end of the screen.
LEAK RATE
COR
•
This default, althought minor, is blocking. It is necessary to launch a
calibration to make it disappear.
Major fault
•
Warning:
- on the digital display, “Er” is permanently displayed.
- a flashing message occurs on the LCD
LEAK RATE
COR
he digital voice advises the operator of the procedure to follow.
•
Result:
May prevent the leak detector from making a vacuum test or an
autocalibration
•
Message:
To read the messsage
NEXT
Note: A major fault can behave like a temporary minor fault if the
origin of the error has disappeared.
Critical failure
•
Warning:
- on the digital display, “Er” is permanently displayed. All indicators
are turned off.
- on the LCD the clear message of a critical failure is directly
displayed. Details are displayed on line 2 and 3.
LEAK RATE
COR
The digital voice advises the operator of the procedure to follow.
•
Result:
Complete shut down of the leak detector is required.
•
Remedy:
Need the servicing of the leak detector before starting it again.
Содержание ASM 142
Страница 1: ...ASM 142 142 D ASM GRAPH GRAPH D GRAPH D HELIUM LEAK DETECTOR User s Manual ...
Страница 98: ......
Страница 106: ......
Страница 111: ......
Страница 119: ......
Страница 124: ......
Страница 129: ......
Страница 140: ......
Страница 144: ......
Страница 166: ......
Страница 315: ...G 201 Wiring diagrams ASM 142 D index 1 2 ...