2/2
C 510
Bargraph zoom
blin
kin
g
le
d
=
rej
ec
t poi
nt
(1
+
1)
d
ec
ades
1
dec
ad
e
G
B
01265
-
E
d
it
io
n
03
-
Ma
rc
h
04
Analog display
When a bargraph zoom is activated, use the helium signal zero scale:
a blinking led indicates the reject point.
Remote control
& control panel
10
-2
10
-3
10
-12
10
-11
10
-10
10
-9
10
-8
10
-7
10
-6
10
-5
10
-4
10
-3
10
-2
COR
ZERO
10
-4
10
-5
blinking led = reject point
(1 + 1) decades
1 decade
INLET
10
3
10
2
10
1
1
10
-1
10
-2
TEST
HIGH
SNIF
VENT
10
-6
10
-7
10
-8
10
-9
10
-10
10
-11
Control panel
The helium signal zero scale displays
the leak value in 2 colors following
the measured leak value:
10
-3
10
-12
mbar
mbar.l/s
CYCLE
AUTO CAL
ZERO
•
if the measured leak value is under
the reject point, the flashing leds are
green,
•
if the measured leak value exceeds
the reject point, the flashing leds are
red (and the blinking led orange).
Reject point
C 211
Remote control
Zero function &
Bargraph zoom
When bargraph zoom and zero functions are ON in the same time,
the operator must read measured leak value in this way as follow:
•
digital display
The leak value displayed is the value corrected with zero function.
Zero function
C 540
•
analog display
Use the helium signal zero scale.
The analog display is the actual bargraph zoom display (see above).
Содержание ASM 142
Страница 1: ...ASM 142 142 D ASM GRAPH GRAPH D GRAPH D HELIUM LEAK DETECTOR User s Manual ...
Страница 98: ......
Страница 106: ......
Страница 111: ......
Страница 119: ......
Страница 124: ......
Страница 129: ......
Страница 140: ......
Страница 144: ......
Страница 166: ......
Страница 315: ...G 201 Wiring diagrams ASM 142 D index 1 2 ...