C 302
Advanced internal calibration of the leak
detector
GB
00970
-
E
d
it
io
n
05
-
J
une
08
Checking function
Internal calibration activated (on), checking function performs a
calibration checking depending on set parameters.
The calibration checking is performed with the internal calibrated leak
of the leak detector (position parameter = internal).
C 305
On request calibration
checking
At any time, the operator could perform a calibration checking with
the internal calibrated leak.
STDBY
+
Time between
both < 5 s.
The leak detector compares the signal delivered by the internal
calibrated leak to the internal calibrated leak parameters set:
•
if the difference is lower than 15 %, the calibration of the leak
detector is OK.
•
if the difference is higher than 15 %, there is a warning message
which requests to perform a complete calibration of the leak detector.
C 301
Parameters setting
Adjustment
C 110
Automatic
Operator
Adjustment
Adjustment
Cycles counter
Hours counter
If checking function is automatic, operator should adjust cycles and
hours counters which will trigger the automatic calibration checking.
Note: The first of the 2 counters reached will trigger the automatic
checking.
Tickets If a printer is connected to the leak detector, a ticket is automatically
printed at the end of the calibration: example
5 .
B 320
2/3
Содержание ASM 142
Страница 1: ...ASM 142 142 D ASM GRAPH GRAPH D GRAPH D HELIUM LEAK DETECTOR User s Manual ...
Страница 98: ......
Страница 106: ......
Страница 111: ......
Страница 119: ......
Страница 124: ......
Страница 129: ......
Страница 140: ......
Страница 144: ......
Страница 166: ......
Страница 315: ...G 201 Wiring diagrams ASM 142 D index 1 2 ...