A help for control panel utilization/access.
Operating principle of the control panel
C 110
Access to level
4
- Password
C 130
Setting and maintenance part presentation of the control panel
Access to parameters and parameters active
depending on authorization
C 120
Summary of screens
Complete displays list with access
way and associated sheet
C 140
Purpose
The helium pollution prevention function prevents the detector from
being polluted by helium when the part or installation to be tested
contains a lot of helium.
Leak detector in test mode, if the signal increases quickly above
depollution reject point, the cycle is automatically ended and the leak
detector stays in stand-by mode until the helium has decreased.
Note: There is no message to inform the operator that the detector is
over again ready for a new test.
Note:
Helium pollution prevention
function not activated
+
Big quantity of Helium in
the pumped parts
=
Detector polluted
Depollution
= Wait and pump a few hours (the best
solution is one night)
= Purge with N2 flow or air flow
Note: When the detector reaches the 10
-4
mbar threshold, it
automatically pass in gross leak test mode.
Activate/Deactivate
the Helium pollution
prevention
2x
Adjustment
Adjustment
C 110
off
on
Содержание ASM 142
Страница 1: ...ASM 142 142 D ASM GRAPH GRAPH D GRAPH D HELIUM LEAK DETECTOR User s Manual ...
Страница 98: ......
Страница 106: ......
Страница 111: ......
Страница 119: ......
Страница 124: ......
Страница 129: ......
Страница 140: ......
Страница 144: ......
Страница 166: ......
Страница 315: ...G 201 Wiring diagrams ASM 142 D index 1 2 ...