G 400
Long distance sniffer probe user manual
GB
01962
-
E
d
it
io
n
03
-
J
une
08
Needle replacement
(continued)
With needlenose pliers, take out the original needle.
Put the new needle (ref.
1
) and cut to the same length as the original
needle.
Note:
For the flexible sniffer probe (SNCxE3Tx and SNCxE4Tx), the needle
length should be 2.5 cm/0.98 inch (ref.
2
).
Bend the needle as shown in the picture
3
.
Put the needle in the nipple, the crooked end to outside (ref.
4
) and
push it in with a 2.5 allen key to stop.
Put back the filter.
Needle
Needle for
flexible sniffer probe
1
2
3
4
Key
O'ring
installation
When you unscrew the nipple from the probe,
it is possible for the o’ring to come out with the
nipple : you should put it back (ref.
1
).
Place the o’ring on the nipple of the aspiration
tube in the sniffer body (ref.
2
).
Push the o’ring with the nipple of the sniffer
probe (ref.
3
).
1
Nipple
O'ring
Body
The o’ring is correctly placed on the nipple (ref.
4
).
Screw on the nipple of the sniffer probe.
2
3
4
7/8
Содержание ASM 142
Страница 1: ...ASM 142 142 D ASM GRAPH GRAPH D GRAPH D HELIUM LEAK DETECTOR User s Manual ...
Страница 98: ......
Страница 106: ......
Страница 111: ......
Страница 119: ......
Страница 124: ......
Страница 129: ......
Страница 140: ......
Страница 144: ......
Страница 166: ......
Страница 315: ...G 201 Wiring diagrams ASM 142 D index 1 2 ...