A help for control panel utilization/access.
Operating principle of the control panel
C 110
Access to level
4
- Password
C 130
Setting and maintenance part presentation of the control panel
Access to parameters and parameters active
depending on authorization
C 120
Summary of screens
Complete displays list with access
way and associated sheet
C 140
Definition
The correction factor is used to display the helium signal (changing
unit for example).
The correction factors applied to the digital display with respect to the
external calibration are:
•
VACUUM COR in gross leak test mode,
•
SNIF COR in sniffing test mode.
The VACUUM/SNIF COR values are automatically adjusted according
to helium signal fluctuations.
These correction factors are memorized until another external
calibration is validated.
They can be activated, deactivated or modified.
The correction factors are automatically calculated by the external
calibration but it is also possible to enter them manually
C 303
Activate/Deactivate
the correction factor
VACUUM/SNIF COR
Adjustment
There are 2 possible procedures depending on which authorized
level: level
3
and
4
.
Procedure with user
level
3
Adjustment
C 110
on
off
VACUUM COR
Adjustment
SNIF COR
Содержание ASM 142
Страница 1: ...ASM 142 142 D ASM GRAPH GRAPH D GRAPH D HELIUM LEAK DETECTOR User s Manual ...
Страница 98: ......
Страница 106: ......
Страница 111: ......
Страница 119: ......
Страница 124: ......
Страница 129: ......
Страница 140: ......
Страница 144: ......
Страница 166: ......
Страница 315: ...G 201 Wiring diagrams ASM 142 D index 1 2 ...