E 400
Basic maintenance of the analyzer cell
GB
00046
-
E
d
it
io
n
07
-
J
une
08
Tighten all
the connection screws
g
Make sure that all the connection screws are facing inwards and
are not in contact with each other.
INCORRECT
CORRECT
Screws facing
outwards: risks of
contact with cell
body
Screws
facing
inwards
Top view
Correct
Incorrect
Reassemble the flange
g
Reposition the special elastomer seal on the body. If it is damaged,
change it.
5/8