E 400
Basic maintenance of the analyzer cell
GB
00046
-
E
d
it
io
n
07
-
J
une
08
The frequencies of preventive maintenance tasks are
listed in section
D 100.
Components
F 500
Safety instructions
B 100
Special
precautions
CAUTION
Disconnect the detector from the main power.
The analyzer cell is very sensitive to any form of
contamination, particularly to dust and electrostatic discharge.
When assembling, to avoid gettering due to dust or finger prints, you
are advised to work:
in a clean room,
on lent free paper,
with unpowered vinyl gloves (clean room gloves),
to blow off each part with filtered dry air,
to block all the openings in the vacuum lines and the VHS
preamplifier.
The operator must take all necessary measures to avoid transferring
electrostatic charges during the operation.