D 100
Table of preventive maintenance intervals
ASM
102
S
ASM
122
D
ASM
142
ASM
142
S
ASM
142
D
ASM
G
ra
ph
D+
ASM
182
T
ASM
192
T
ASM
192
T
2
ASM
182
T
D+
ASM
192
T
D+
ASM
192
T
2D
+
ASM
1002
ASI
22
1000 h
(1)
Clean filters (inlet filters, air inlet
filter)
-
•
•
•
•
•
•
•
2000 h
(1)
or
3 months
(2)
Change the rotary vane pump oil.
Change the oil mist eliminator.
E 750
•
•
•
•
•
•
4000 h
(1)
or
6 months
(2)
Clean the vacuum lines, the valves
and the gauges with alcohol - Dust
the electronic boards and the fans.
Partial maintenance of the analyzer
cell: Replace analyzer cell
filaments and collector. Clean the
analyzer cell with alcohol (this
cleaning may be necessary in case
of general internal contamination
creating insulating deposits).
E 400
•
•
•
•
•
•
•
•
•
•
•
•
•
•
8000 h
(1)
ou 1 year
(2)
Sniffer probe filter replacement if
used.
G 200
•
•
•
•
•
•
•
•
•
•
•
•
•
Pirani gauge adjustment.
CS
•
•
•
•
•
APT 1004 gauge adjustment.
CS
•
•
Electronic overhaul
-
•
•
Replace the seal in the rotary vane
pump.
CS
•
•
•
AMD1 pump: replacement of
membranes and check valves.
Change the inlet filter of the test
chamber.
E 710
•
GB
02469
-
E
d
it
io
n
03
-
J
une
08
Frequency* Maintenance operations to be performed
CS : Please contact Customer Service
*Service intervals: The service intervals given are for applications and work rates which conform to the normal operating
conditions. If the machine is operating under more difficult conditions they can be shortened.
(1) running time
(2) running time or storage
(3) storage
1/2
Содержание ASM 142
Страница 1: ...ASM 142 142 D ASM GRAPH GRAPH D GRAPH D HELIUM LEAK DETECTOR User s Manual ...
Страница 98: ......
Страница 106: ......
Страница 111: ......
Страница 119: ......
Страница 124: ......
Страница 129: ......
Страница 140: ......
Страница 144: ......
Страница 166: ......
Страница 315: ...G 201 Wiring diagrams ASM 142 D index 1 2 ...