ECG Monitoring and Pacemakers
9650-001355-01 Rev. M
X Series Operator’s Guide
6-9
ECG Monitoring and Pacemakers
When the unit performs ECG monitoring on a patient with an implantable pacemaker, the unit’s
Pacer indicator feature can indicate the occurrence of pacemaker signals.
If the Pacer Indicator setting is
ON
, the X Series performs the following actions:
•
detects the implantable pacemaker pulses
•
blanks the pacemaker pulses from the waveform—preventing them from disturbing the
ECG waveform and allowing for an accurate QRS detection
•
displays and prints vertical dashed lines to indicate the detected pacemaker signals
If the Pacer Indicator setting is
OFF
, the X Series does not perform the following actions:
•
detect the implantable pacemaker pulses
•
blank the pacemaker pulses from the waveform
•
display or print the vertical dashed line pacer markers
You can turn the Pacer Indicator
ON/OFF
from the Setup>ECG menu.
To turn the Pacer Indicator
ON/OFF
:
1. Press the
More
quick access key (
).
2.
Press the
Setup
quick access key (
).
3.
Use the navigation keys to select
ECG
then press
.
4.
Use the navigation keys to highlight and select
Patient Pace Marker
then press
.
5.
Use the navigation keys to select
On
or
Off
as appropriate. When you select
Off
, the Pacer
Off Indicator(
) icon appears in the top ECG trace on the Display Screen to indicate the
pacer indicator is off. If the patient has a cardiac pacemaker, the Paced Indicator should be
set to
On
indicating that pace pulse detection is On.
There are situations where ECG artifact could simulate pacemaker signals which could cause
false pacemaker detection and blanking. These situations may cause inaccurate QRS detection
and it may be desirable to turn the Pacer Indicator
off
. Inversely, when the Pacer Indicator
setting is
OFF
, implantable pacemaker signals may cause inaccurate QRS detection and it may
be desirable to turn the Pacer Indicator on.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...