CO2 System Messages
9650-001355-01 Rev. M
X Series Operator’s Guide
9-11
CO
2
System Messages
When monitoring CO
2
, the X Series unit may display the following messages:
System Message
Cause
INITIALIZING
The unit is initializing the CO
2
monitoring function
measurement, is functioning normally, and will display
CO
2
measurement after a brief delay.
CHECK FILTERLINE
The sampling line is not connected. Check sampling line
connection. Replace airway adaptor or cannula, if
defective.
FILTERLINE OCCLUSION
The sampling or exhaust line is blocked. Check sampling
and exhaust lines. Make sure that the sampling line and
any inputs to the patient breathing apparatus are not
connected to the exhaust port. If the current FilterLine is
correctly attached, replace the FilterLine.
AUTO ZERO
The X Series unit displays
AUTO ZERO
when the CO
2
module performs periodic self-maintenance. During self-
maintenance, the CO
2
module performs one or more of
the following tests: ambient pressure measurement, auto
zero, and a flow test. The CO
2
module completes its self-
maintenance tests in approximately 10 seconds.
PURGING
The CO
2
module performs a purging whenever it detects
an occlusion in the line or airway adaptor, and displays
PURGING
while it tries to clear the occlusion. If the CO
2
module cannot clear the occlusion within 30 seconds, the
X Series unit issues an equipment alert and displays the
message
FILTERLINE OCCLUSION.
At this point, if the
current FilterLine is correctly attached, replace the
FilterLine.
VALUE OVER RANGE
The CO
2
value exceeds the specified range.
CO2 DISABLED
An error has occurred with the CO
2
module, and the X
Series unit has disabled the module for the duration of the
unit’s power cycle. If the problem persists, the unit may
require servicing.
CO2 MONITORING FAILURE
The X Series unit has determined that the CO
2
module is
not functioning correctly. If the problem persists, the unit
may require servicing.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...