Chapter 10 Pulse CO-Oximetry (SpO2)
10-4
www.zoll.com
9650-001355-01 Rev. M
•
Inaccurate SpHb and SpOC readings may be caused by:
•
Improper sensor application
•
Intravascular dyes such as indocyanine green or methylene blue
•
Externally applied coloring and texture such as nail polish, acrylic nails, glitter, etc.
•
Elevated PaO2 levels
•
Elevated levels of bilirubin
•
Low arterial perfusion
•
Motion artifact
•
Low arterial oxygen saturation levels
•
Elevated carboxyhemoglobin levels
•
Elevated methemoglobin levels
•
Hemoglobinopathies and synthesis disorders such as thalassemias, Hb s, Hb c, sickle
cell, etcetera
•
Vasospastic disease such as Raynaud's
•
Elevated altitude
•
Peripheral vascular disease
•
Liver disease
•
EMI radiation interference
•
Inaccurate SpO2 readings may be caused by:
•
Improper sensor application.
•
Elevated levels of COHb or MetHb: High levels of COHb or MetHb may occur with a
seemingly normal SpO2. When elevated levels of COHb or MetHb are suspected,
laboratory analysis (co-Oximetry) of a blood sample should be performed.
•
Intravascular dyes, such as indocyanine green or methylene blue.
•
Externally applied coloring and texture, such as nail polish, acrylic nails, glitter, etc.
•
Elevated levels of bilirubin
•
Severe anemia
•
Low arterial perfusion
•
Motion artifact
•
Interfering Substances: Dyes or any substance containing dyes that change usual blood
pigmentation may cause erroneous readings.
•
The pulse co-oximeter is intended only as an adjunct device in patient assessment. It should
not be used as the sole basis for diagnosis or therapy decisions. It must be used in
conjunction with clinical signs and symptoms.
•
The pulse co-oximeter is not an apnea monitor.
•
The pulse co-oximeter may be used during defibrillation, but this may affect the accuracy or
availability of the parameters and measurements.
•
The pulse co-oximeter may be used during electrocautery, but this may affect the accuracy
or availability of the parameters and measurements.
•
The pulse co-oximeter should not be used for arrhythmia analysis
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...