Sending Disclosure Logs
9650-001355-01 Rev. M
X Series Operator’s Guide
22-13
Sending Disclosure Logs
For systems with a Disclosure log server configured, the X Series unit allows you to send
patient disclosure logs for up to 15 cases at one time to a remote server through a wireless
connection.
Note:
The X Series unit automatically cancels any case disclosure log transmissions when
defibrillation is activated.
To transfer disclosure log data:
1. From the Primary Display window press the
Log
button (
) then press the
Disclosure
Log Transfer
quick access key (
). The Close Current Case? dialog box displays.
2.
Specify whether or not to include the current case in the transfer by navigating to and
selecting:
•
Close Case
if you would like to include the disclosure logs for the current case in the
transfer. When you select
Close Case
, the X Series closes the current case and generates a
new case for the current patient. That way Pacing, Defib, and AED parameter settings for
the current patient remain intact following the transfer.
•
Continue
button to continue the transfer without the current case.
The
Case Selection
dialog box displays with a list of up to 150 most recent cases.
3.
Use the navigation keys to select
Select Cases
then press the
Select
button (
)
.
The
highlight moves to the list of available cases to transfer.
4.
Select up to 15 cases to include in the transfer by using the up and down navigation arrows
to highlight a case then pressing the
Select
button (
). A case is selected for inclusion in the
disclosure log transfer when a check mark displays to the left of the case. The
Number of
Cases Selected
field indicates the number of cases currently selected.
5.
When you are done selecting cases, press the
Home/Display
button (
). The highlight
moves to the
Transfer
key.
6.
With the
Transfer
key highlighted, press the
Select
button (
). The Disclosure Log
Transfer dialog displays indicating the transfer progress.
If the Disclosure Log Transfer Failure dialog box displays, use the navigation keys to select
Retry
then press the
Select
button (
). The X Series continues the transfer from the case
that failed. Otherwise, if you would like to end the transfer use the navigation keys to select
Cancel
then press the
Select
button (
) to return to the Primary Display window.
7.
When the Disclosure Log Transfer Complete dialog box displays select
OK
to return to the
Primary Display window.
LOG
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...