Chapter 24 Maintenance
24-10
www.zoll.com
9650-001355-01 Rev. M
Signature
_______________________________________________
X Series OPERATOR’S SHIFT CHECKLIST
Date: ________________ Shift: _______________ Location: _______________
Mfr/Model No.: ____________________________ Serial No. or Facility ID No.: _________________
At the beginning of each shift, inspect the unit. Indicate whether all requirements have been met.
Note any corrective actions taken. Sign the form.
Okay as found
Corrective Action/Remarks
1. Defibrillator Unit
Clean, no spills, and casing undamaged
2. Cables/Connectors
a. Inspect for damaged insulation, frayed/broken wires, or bent
connector pins
b. Connectors engage securely
3. Sensors
(pulse oximetry, NIBP cuff and hose, temperature sensors)
a. Inspect for signs of damage or excessive wear
b. Connectors engage securely
4. Paddles
a. Clean, not pitted, or damaged.
b. Switches operate freely
c. Cables and connectors free of damage and engage securely
5. Supplies and Accessories
a. Therapy pads in sealed pouches (2 sets, not expired)
b. Defib gel or gel pads
c. ECG monitoring electrodes
d. Alcohol wipes
e. Razors/scissors
f. Recorder paper
6. Batteries
a. Fully charged battery installed in unit
b. Fully charged spare battery available
7.
Ready For Use Indicator shows Ready -
No flashing or Do Not Use symbol
8. Operational Checks
A. Power Up Sequence
a. Battery charge and AC indicators illuminate with AC power
b. Audio beeps heard and Visual Alarm indicators briefly illuminate
c. Self Test passed
B. Hands Free Defibrillation (Test with only battery power)
a.
CHECK THERAPY ELECTRODES
and
PADS SHORT
DETECTED
messages display
b. Charge time < 7 seconds
c.
DEFIB SHORT TEST
passed at 30 Joules
C. Pacer Test (Test with only battery power)
a. Printer prints Pace markers every 25 mm at 60 ppm
b.
PACING: SHORT DETECTED
message displays-- no error
messages at 100 mA
c.
PACING: CHECK THERAPY ELECTRODES
message displays
D. Paddles
a.
APPLY PADDLES TO PATIENT
message displays when paddles
are connected to MFC
b. Paddle switches functional (Recorder, Energy Select, Charge, Shock)
c.
DEFIB SHORT TEST
passed at 30 Joules
E. Reconnect Unit to AC Power
____________ Major problem(s) identified (OUT OF SERVICE)
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...