![ZOLL X Series Operator'S Manual Download Page 26](http://html1.mh-extra.com/html/zoll/x-series/x-series_operators-manual_4055648026.webp)
Chapter 1 General Information
1-14
www.zoll.com
9650-001355-01 Rev. M
Ready For Use (RFU) Indicator
The X Series has an RFU indicator on the front panel that indicates if the device is ready for
use. The RFU indicator has three states which are described in the following table.
State
Description
Action
Ready for Use
The device is ready for use. Patient
monitoring, defibrillation, and
pacing parameters are functional
and the battery is above the low
battery capacity.
Note
: If the device is plugged into
the auxiliary power adapter, the
Ready for Use indicator may display
even if the battery is depleted.
Check the status of the battery
before removing the device from the
auxiliary power adapter.
None required.
Flashing
One or more of the following has
occurred:
•
The battery is not properly
installed.
•
A low battery is installed.
•
A battery fault has occurred.
•
There is no battery installed
while connected to auxiliary
power.
•
One or more patient monitoring
parameters have failed self-test
(NIBP, SpO
2
, CO
2
, IBP, or
Temp).
•
The front panel button self-test
failed.
•
The speech database self-test
failed.
Install a fully charged battery in the
unit and check the RFU indicator
again. If the RFU indicator
continues to flash, remove the unit
from service and contact the
appropriate technical personnel or
the ZOLL Technical Service
Department.
Do Not Use
One or more of the following has
occurred:
•
The battery is not properly
installed.
•
No battery is installed and
auxiliary power is not present.
•
A very low battery (below
software shutdown limit) was
installed.
•
ECG, defibrillator, or pacer self-
tests have failed, or other critical
self-tests have failed.
Install a fully charged battery in the
unit and check the RFU indicator
again. If the RFU indicator
continues to display the Do Not
Use symbol, remove the unit from
service and contact the appropriate
technical personnel or the ZOLL
Technical Service Department.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...