Chapter 24 Maintenance
24-2
www.zoll.com
9650-001355-01 Rev. M
Daily/Shift Check Procedure
Inspection
Equipment and Accessories
•
Ensure that the X Series unit is clean (with no fluid spills) and free of visible damage.
•
Inspect all cables, cords, and connectors for signs of damage or excessive wear (cuts in
insulation, fraying, broken wires, dirty or bent connector pins). Replace if damaged.
•
Inspect the battery, ECG patient cable, multifunction cable (MFC), OneStep cable, pulse
oximetry sensor, blood pressure cuff, and hose and temperature sensors for signs of damage
or excessive wear. Replace if damaged.
•
Inspect the defibrillator paddles for damage or excessive wear. Check that paddle surfaces
are clean and free of electrolyte gel or other contaminants. Verify that all paddle controls
operate freely when pressed and released.
•
Inspect the external AC adapter and its power cords for signs of damage. Replace if
damaged.
Supplies
•
Verify the presence, proper condition, and appropriate quantities of all disposable supplies
(hands-free therapy electrodes, ECG monitoring electrodes, defibrillator gel, recorder paper,
alcohol swabs, razors/scissors, etc.).
•
Verify that two sets of hands-free therapy pads are available.
•
Ensure that the therapy and monitoring electrodes are sealed within the packages and within
the expiration dates printed on the packages.
•
Open the recorder door on the left side of the X Series unit and verify that an adequate
supply of paper is available in the unit.
Batteries
•
Verify that a charged battery is fully inserted into the battery compartment of the X Series
unit.
•
Ensure that a fully charged spare battery pack is available.
Ready For Use Indicator
•
Check the Ready For Use (RFU) indicator on the X Series front panel. Verify that the RFU
symbol is displayed.
If the RFU indicator is flashing or displaying the Do Not Use symbol, then either an
equipment fault has been detected or the unit's battery is low or not properly installed. Install
a fully charged battery in the unit and check the RFU indicator again. If the RFU indicator
continues to flash or display the Do Not Use symbol, remove the unit from service and
contact the appropriate technical personnel or the ZOLL Technical Service Department.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...