Chapter 7 Monitoring Respiration (Resp) and Heart Rate (HR)
7-2
www.zoll.com
9650-001355-01 Rev. M
Respiration/Breath Rate Meter
If enabled, the X Series unit displays the patient’s respiration in the Respiration/Breath Rate
Meter.
The respiration
meter displays the respiration rate that it derives, by default, from the unit’s
CO
2
monitoring function. If CO
2
monitoring is not available, the unit derives the respiration
rate through
impedance pneumography
, using a specified ECG electrode configuration. If ECG
monitoring isn’t functioning, the
Resp/BR
meter will not display a respiration rate.
Using Impedance Pneumography to Measure Respiration
Impedance pneumography detects respiration by applying a high-frequency, low-current AC
signal to the patient and measuring the changes in impedance through ECG electrode Lead
l
(RA-LA) or Lead
ll
(RA-LL). As the patient inhales and chest volume expands, impedance
increases; as the patient exhales, impedance decreases.
Warning!
•
Impedance pneumography detects respiratory effort through changes in chest
volume. However, No Breath episodes with continued respiratory effort may go
undetected. Always monitor and set alarms for SpO
2
when using impedance
pneumography to monitor respiratory function.
•
With any monitor that detects respiratory effort through impedance pneumography,
artifact due to patient motion, apnea mattress shaking, or electrocautery use may
cause apnea episodes to go undetected. Always monitor and set alarms for SpO
2
when using impedance pneumography to monitor respiratory function.
•
When using impedance pneumography, don’t use the X Series unit with another
respiration monitor on the same patient, because the respiration measurement
signals may interfere with one another.
•
Impedance pneumography is
not
recommended for use on paced patients, because
pacemaker pulses may be falsely counted as breaths.
•
Impedance pneumography is not recommended for use with high frequency
ventilation.
•
Since impedance pneumography uses the same leads as the ECG channel, the
X Series unit determines which signals are cardiovascular artifact and which signals
are the result of respiratory effort. If the breath rate is within five percent of the
heart rate, the monitor may ignore breaths and trigger a respiration alarm.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...