Analysis/CPR Protocol Defibrillation Procedure
9650-001355-01 Rev. M
X Series Operator’s Guide
17-5
When you press the Energy Select Up/Down arrow keys or
CHARGE
button while in
Analysis/CPR Protocol mode, the unit transitions to Manual mode. If the unit is charged when
you transition from Analysis/CPR Protocol to Manual mode, the unit disarms and halts any
analysis that is taking place.
Non-Shockable Rhythm
When a nonshockable rhythm is detected, the unit displays a
NO SHOCK ADVISED
message.
The unit then leads you through a CPR interval and then restarts ECG analysis automatically.
The Analysis/CPR Protocol mode repeats the Analysis and CPR intervals as long as Analysis/
CPR Protocol is active. You can press the
Exit
quick access key at any time to return to Manual
mode.
Note:
When a nonshockable rhythm is detected, the X Series does not prevent the user from
manually defibrillating the patient.
Shockable Rhythm
If the patient’s rhythm is shockable, the unit displays the
SHOCK ADVISED
and
PRESS
SHOCK
messages. The defibrillator automatically prompts the operator to shock the patient at
the preconfigured energy level and the
SHOCK
button illuminates.
A continuous tone sounds for 20 or 50 seconds (depending on configuration), followed by a
higher pitch tone for 10 seconds.You must deliver the shock within this 30 or 60 second interval
(depending on configuration), or the defibrillator will disarm itself
3 Press SHOCK button
WARNING! Warn all persons in attendance of the patient to STAND CLEAR prior to defibrillator
discharge.
Do not touch the bed, patient, or any equipment connected to the patient during
defibrillation. A severe shock can result. Do not allow exposed portions of the patient’s
body to come in contact with metal objects, such as a bed frame, as unwanted pathways
for defibrillation current may result.
Press and hold the illuminated
SHOCK
button
on the front panel until energy is delivered
to the patient.
The delivered energy level is displayed at the bottom of the screen and the shock number (1)
displays at the top of the screen and in the Defib Control panel at the bottom of the screen.
The unit then leads you through a CPR interval and then restarts ECG analysis automatically.
The Analysis/CPR Protocol mode repeats the Analysis and CPR intervals as long as Analysis/
CPR Protocol is active. You can press the
Exit
quick access key at any time to return to Manual
mode.
1
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...