Chapter 8 Monitoring Non-Invasive Blood Pressure (NIBP)
8-10
www.zoll.com
9650-001355-01 Rev. M
Ensuring Correct Cuff Inflation Settings
Before taking an NIBP measurement for a new patient, ensure that the cuff inflation settings are
appropriate for that patient.
Check that the correct patient mode is selected. The initial Cuff Inflation Target (CIT) pressure
is dependent on the patient type and the configured CIT preset. The default and configurable
CIT presets for each patient type are indicated in the following table (default values are in
bold).
Note:
You can configure the CIT pressure from the Setup>NIBP menu.
In order to accurately measure systolic pressure, the CIT pressure must be high enough to
occlude the underlying artery. However, setting the CIT pressure too high may unnecessarily
increase the reading determination time and patient discomfort.
After each NIBP measurement, the X Series unit adjusts the cuff inflation pressure to optimize
the next NIBP measurement.
Note:
The maximum cuff inflation pressure for neonates is 153 mmHg.
Warning!
Before using the X Series unit to monitor a new patient, power down the unit for
at least
2 minutes
to reset all patient parameters and eliminate all adjustments made for the
previous patient.
Adult
Pediatric
Neonate
120 mmHg
80 mmHg
60 mmHg
140 mmHg
90 mmHg
70 mmHg
160 mmHg
100 mmHg
80 mmHg
180 mmHg
110 mmHg
90 mmHg
200 mmHg
120 mmHg
100 mmHg
220 mmHg
130 mmHg
110 mmHg
240 mmHg
140 mmHg
120 mmHg
260 mmHg
150 mmHg
130 mmHg
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...