Warnings
9650-001355-01 Rev. M
X Series Operator’s Guide
1-17
ECG Monitoring
•
Implanted pacemakers might cause the heart rate meter to count the pacemaker rate during
incidents of cardiac arrest or other arrhythmias. Dedicated pacemaker detection circuitry
may not detect all implanted pacemaker spikes. Check the patient's pulse; do not rely solely
on heart rate meters. Patient history and physical examination are important factors in
determining the presence of an implanted pacemaker. Pacemaker patients should be
carefully observed. See “Pacemaker Pulse Rejection:” on page A-15 of this manual for
disclosure of the pacemaker pulse rejection capability of this instrument.
•
Use only ECG electrodes that meet the AAMI standard for electrode performance
(AAMI EC-12). Use of electrodes not meeting this AAMI standard could cause the ECG
trace recovery after defibrillation to be significantly delayed.
•
Prior to attempting synchronized cardioversion, ensure the ECG signal quality is good and
that sync markers are displayed above each QRS complex.
•
Do not place electrodes directly over an implanted pacemaker.
•
The X Series unit detects ECG electrical signals only. It does not detect a pulse (effective
circulatory perfusion). Always verify pulse and heart rate by physical assessment of the
patient. Never assume that the display of a nonzero heart rate means that the patient has a
pulse.
•
Excessive artifact can result due to improper skin preparation of the electrode sites. Follow
skin preparation instructions in Chapter 6: “Monitoring ECG.”
•
Equipment such as electrocautery or diathermy equipment, RFID readers, electronic article
surveillance (EAS) systems, or metal detectors that emit strong radio frequency signals can
cause electrical interference and distort the ECG signal displayed by the monitor, thereby
preventing accurate rhythm analysis. Ensure adequate separation between such emitters, the
device, and the patient when performing rhythm analysis.
•
Shock Hazard: Use of accessories, other than those specified in the operating instructions,
may adversely affect patient leakage currents.
•
Certain line-isolation monitors may cause interference on the ECG display and may inhibit
heart rate alarms.
•
Monitoring ECG through the paddles may result in inaccurate heart rate display due to
artifact.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...