CPR Voice Prompts (Adult Only)
9650-001355-01 Rev. M
X Series Operator’s Guide
19-3
CPR Voice Prompts (Adult Only)
The X Series unit can be configured to issue voice prompts related to the depth of chest
compressions as feedback to rescuers performing CPR. Two voice prompts are available for
this purpose:
•
Push Harder
•
Good Compressions
When chest compressions are detected but their depth is consistently less than the target depth
(AHA/ERC 2005: 1.75, AHA/ERC 2010: 2.0), the device periodically issues the “Push Harder”
voice prompt. If the rescuer responds by increasing compression depth to more than the target
depth on a consistent basis, the unit issues a “Good Compressions” prompt.
CPR voice prompts are only available when using Adult CPR electrodes.
CPR Metronome
The X Series unit includes a CPR metronome feature that can be used to encourage rescuers to
perform chest compressions at the AHA/ERC recommended rate of 100 compressions per
minute.
The CPR metronome can be configured to operate in all modes (AED, Manual, and Analysis/
CPR Protocol), or in AED mode only. The CPR metronome can also be disabled for all modes.
When activated, the metronome beeps at the AHA/ERC recommended rate to provide a
compression rhythm for rescuers to follow.
The CPR metronome’s Constant Metronome feature allows the metronome to continue beeping
under certain circumstances. When Enable Metronome is set to Never in the Supervisor>CPR
menu, Constant Metronome is greyed out.
The CPR metronome can be configured as follows.
Constant Metronome Enabled
AED Mode
•
If Enable Metronome is set to Always or AED Mode Only, the metronome is active when
chest compressions are first detected during the CPR interval. The metronome is silent when
the CPR idle timer starts.
•
If set to Never, the metronome is silent.
Analysis/CPR Protocol or Manual Mode
•
If Enable Metronome is set to Always, the metronome is active when chest compressions
are first detected during the CPR interval. The metronome is silent when the CPR idle timer
starts.
•
If set to AED Mode Only or Never, the metronome is silent.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...