![background image](http://html1.mh-extra.com/html/zoll/x-series/x-series_operators-manual_4055648169.webp)
Enabling/Disabling Temperature Alarms and Setting Alarm Limits
9650-001355-01 Rev. M
X Series Operator’s Guide
12-3
Enabling/Disabling Temperature Alarms and Setting Alarm
Limits
When enabled, the X Series unit sounds alarms whenever temperature measurements are
outside set limits.
You can enable (or disable) temperature alarms and set the upper and lower alarm limits
through the Alarms quick access key or the Temp
Parameter Control Panel.
To configure temperature alarms through the
Alarms
quick access key:
1. Press the More
quick access key (
).
2.
Press .
3.
Press the
Limits
quick access key (
).
4.
Use the navigation keys to highlight and select the
T1 Alarm
,
T2 Alarm
, or
T Alarm
menu selection. On the Alarm Settings menu, use the navigation keys to select the fields that
you want to change. The fields are
•
Upper Limit Enable
•
Lower Limit Enable
•
Upper Limit
•
Lower Limit
5.
When you are finished changing values on the alarm
menu, navigate to the Backarrow key
to exit the menu.
Setting Upper and Lower Temperature Alarm Limits
Initially, the Temperature Alarm Settings
menu specifies that the Temperature alarms are
enabled or disabled and displays the default upper and lower limits. The upper and lower limits
can be
ON
or
OFF
(default is
OFF
). The following table lists the default Temperature limits for
adult, pediatric, and neonate patients, and gives the range in which you can set these limits:
Limits
Patient Type
Temperature Limit Default
Temperature Limit Range
Adult
Lower: 35
C (95.0
F)
Upper: 37.8
C (100
F)
Lower: 0.0 - 48.0
C (32.0 - 120.0
F)
Upper: 2.0 - 50.0
C (34.0 - 122.0
F)
Pediatric
Lower: 35
C (95.0
F)
Upper: 37.8
C (100
F)
Lower: 0.0 - 48.0
C (32.0 - 120.0
F)
Upper: 2.0 - 50.0
C (34.0 - 122.0
F)
Neonate
Lower: 35
C (95.0
F)
Upper: 37.8
C (100
F)
Lower: 0.0 - 48.0
C (32.0 - 120.0
F)
Upper: 2.0 - 50.0
C (34.0 - 122.0
F)
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...